Pressure Transducers Potentiometric Pressure Strain Gage PressurePressure Transducers Tutorial

 High AC or DC Outputs

GENERAL DESCRIPTION

Sensor Systems LLC Pressure Transducers provide a precise electrical voltage output as a function of applied pressure, with an accuracy previously unattainable. Instruments are available to cover the pressure ranges of 0-0.1 psi through 0-10,000 psi gage, absolute, and differential with linear or non-linear outputs that are tailored specifically for your application.
 
Sensor Systems LLC represents the highest development in the pressure potentiometer field, and while eliminating all the shortcomings of the old-fashioned wirewound types; they retain the basic advantages of simplicity, high voltage output and low cost.
 
State-of-the-art technology includes a unique and proprietary conductive film resistance element that features infinite resolution as well as outstanding life and reliability. Multiple finger wipers directly fastened to the pressure sensing element ride on the mirror-like surface of this resistance element; motion of the pressure-sensing element is thereby converted into a corresponding voltage output. The elimination of intermediate linkages produces minimum sensitivity to vibration, shock, and temperature; additionally, maximum pressure sensitivity, repeatability, reliability and exceptionally long life are assured. Utilizing pre-cycled and pre-stressed pressure sensing element considerably below their allowable stress levels reduces zero shift to a negligible factor.

Why Sensor Systems, LLC Pressure Transducers are the MOST ACCURATE

Sensor Systems LLC was able to substantially eliminate resolution errors by utilizing the special conductive film resistance element. Thus, the output accuracy is solely determined by the linearity of the sensing element (capsule or Bourdon Tube). The resistance element is adjusted to compensate even for this error through the use of a special manufacturing technique outlined below.
 
A schedule of known pressure increments is fed into the sensing element; the voltage output of the transducer is compared in a special bridge circuit with that desired. Through a computer controlled mechanical milling process, the voltage field of the conductive film resistance element is shaped to correspond with the desired voltage ratio. By the same method, linear or non-linear pressure-voltage output relationship not only insures the desired linearity or conformity but also fixes the scale factor of the unit. This allows any one of a lot of Sensor Systems LLC Pressure Transducers to be used interchangeably without further individual calibration or circuit correction by the user. Pressure Switches are likewise precisely adjusted. Various models are available for use in control circuits, indicating instruments, air data computers, or telemetry. True airspeed, Mach, and Thrust data computers are also manufactured to suit.

TECHNICAL DATA

Typical features & performance of Sensor Systems LLC Pressure Transducers includes: 

  • HIGH VOLTAGE OUTPUT
     
  • AC or DC OUTPUT - same as input voltage, insensitive to power supply frequency.
     
  • LINEAR OR FUNCTIONAL OUTPUT - linear, square root, altitude, airspeed, etc.
     
  • SIMPLE INSTALLATION - fixed calibration, insensitive to electrical line lengths.
     
  • LOW COST - requires no auxiliary equipment for signal conditioning, amplifying compensating or converting electrical inputs or outputs.

In all these areas Sensor Systems LLC pressure transducers are superior to LVDT, strain gage, capacitor, semi-conductor and other types of pressure transducers. Sensor Systems LLC pressure transducers have equivalent accuracy and stability; the response to high frequency input pressures is faster than that for the LVDT type and slower than that for the other types.
 
In order to assist you in the selection of the particular pressure transducers described in this catalog, there is discussed below every relevant technical aspect of construction, performance, and use including practical circuitry considerations. While we believe that this material is complete, we will be pleased to personally review with you any aspect of this data or your particular application.
 

 

Fig. 4
Errors due to electrical loading in
Sensor Systems, LLC Transducers
 

Fig. 5
Error bands in transducer output
 

 Pressure Transducer Construction

The pressure devices described in this catalog use an elastic pressure-sensing element acting as a prime mover for positioning an electromechanical transducer, thereby providing an electrical output for a pressure input. In the low pressure ranges the sensing element is a metal capsule consisting of a pair of symmetrical concentrically-corrugated diaphragms, welded together at their outer rims to form a hollow flexible member having predictable motion when fluid or gas pressure applied internally or externally. In the high pressure ranges the sensing element is a metal Bourdon Tube, consisting of a circular tube of oval cross-section, brazed closed at one end, so that pressure applied internally or externally causes the circular tube to partially straighten and its closed tip to move in a predictable path. In all cases the electro-mechanical transducer is a variable resistor, in which the moveable contact or slider is driven by the motion of the capsule or Bourdon Tube. When voltage is applied to the ends of the variable resistor, the moveable contact position appears on the output terminal as a proportionate fraction of the excitation voltage.
 
An Absolute Pressure transducer provides an output voltage which is proportional to the difference between the applied pressure and a perfect vacuum. In such a unit, the pressure-sensing element is completely evacuated and sealed; the Hi pressure port is not present and input pressure is applied through the Lo port.
 
A Gage Pressure transducer provides an output voltage which is proportional to the difference between the applied pressure and the ambient pressure. In such a unit, the input pressure is through the Hi port and the ambient pressure is applied through the open Lo port.
 
A Differential Pressure transducer provides an output voltage which is proportional to the difference between two applied pressures. In such a unit, the higher of the two pressures is applied through the Hi port and the lower through the Lo port.

Where a corrosive or electrically conductive media is used and enters only into the capsule (gage or Hi side of Differential) the transducer construction is as shown in Fig. 1 and Fig. 2; where such a media enters the case, the construction is altered to provide an isolation media for the electro-mechanical transducer. Here the case volume is filled with a neutral fluid and isolated from the Lo port by a slack flexible membrane as shown in Fig. 3. Pressure caused by such media entering the Lo port is transmitted to the pressure-sensing element through the neutral isolation fluid.

The metal alloy and manufacturing process of the pressure capsule or Bourdon Tube determine its corrosive resistance, operating temperature range, temperature sensitivity, life, accuracy, repeatability, hysteresis and long-term stability. A wide selection of alloys and close process control permits an optimization of operating characteristics in the face of a wide range of media and environments. Pressure sensors made of Inconel X, phosphor bronze, beryllium copper or stainless steel are the most corrosion resistant, but generally require internal temperature compensation, in the form of a bi-metallic member, to offset the change in deflection of the sensor resulting from a change in temperature. Pressure sensors made of a nickel alloy called Ni Span "C" are relatively insensitive to temperature and generally require no internal temperature compensation.

Similarly the dimensions of the capsule and Bourdon Tube pressure elements determine the pressure range, vibration and shock sensitivity, as well as the quality of the operating characteristics listed above. In general, larger diameter thinner metal capsules are used at the lower pressures due to their greater flexibility; however such capsules, while more sensitive to input pressure, are also more sensitive to environmental vibration and shock; internal damping or static and dynamic balancing are provided to minimize these effects. High-pressure transducers using either relatively rigid capsules or Bourdon Tubes are less sensitive to these environmental factors. Capsules and Bourdon Tubes are available in a wide variety of sizes and configurations to permit the optimization of transducer operating characteristics.

Likewise, the electromechanical transducer is optimized in the manufacture of each individual transducer through a semi-automatic calibration process to compensate for pressure element non-linearity or errors induced by electrical loading in the output and/or to provide an output which follows any desired variation with input pressure. For example, in the Altitude Transducer, Model 7000, the output is adjusted to be proportional to distance above sea level. In the Model 2000 transducer, the output is adjusted to be linearly proportional to input pressure.
 

 Performance Characteristics of
 Pressure Transducers

Ideally, the output of a pressure transducer can be represented by a perfect straight-line relationship between input pressure and output voltage. In practice, this relationship is closely but not exactly achieved due to various inherent characteristics of the two principal components making up the transducer - the pressure sensing element and the electromechanical transducer.

PRESSURE ELEMENT PERFORMANCE CHARACTERISTICS

The metal pressure-sensing element is an elastic member which, due to the crystalline structure of metals, does not behave as a perfect spring. In order of magnitude, first, its spring rate or the energy per unit deflection is not a constant but varies somewhat with temperature, depending upon the particular metal alloy, and varies also with the total deflection range; these variations in spring rate produce what is called Temperature Error in output in the former case, and Linearity Error in the latter case. Second, the actual deflection at the same pressure is different depending upon whether the pressure is reached in an increasing direction from a lower pressure or in a decreasing direction from a higher pressure; the difference between "upscale" and "down-scale" deflections at the same pressure is called Hysteresis Error. Third, a certain level of input pressure is required to overcome the inherent internal friction of the metal sensing element before an initial deflection takes place. This reflects itself in the absolute deflection as an uncertainty called Resolution Error.

ELECTROMECHANICAL TRANSDUCER PERFORMANCE CHARACTERISTICS

The electromechanical transducer itself does not have a constant output per unit deflection over its entire deflection range, producing a Linearity Error. An effective Linearity Error is also produced by the electrical loading on the resistance element of the electromechanical transducer by the electrical devices outside the pressure transducer to which the moveable slider is connected. The amount of this error is a function of the ratio of resistance element terminal resistance to the effective load resistance. The effective load resistance should be specified when ordering the pressure transducer so that compensation for the loading error can be made during the manufacturing calibration process. Without such compensation the errors in output of Sensor Systems LLC Pressure Transducers due to load ratios greater than 0.001 are shown in Fig. 4. In addition, the moveable contact or slider requires a certain force level from the pressure element to overcome the friction of contact, resulting in a Friction Error. Even though the resistance element in the Sensor Systems LLC electro-mechanical transducer is one continuous smooth-surfaced unit, effects equivalent to Resolution Error and/or Repeatability Error are produced by the finite radius of the moveable contact, which in itself is a spring member.

PRESSURE TRANSDUCER ACCURACY

As reflected in the actual measured output of the pressure transducer in a given environment, the maximum effect of all of these errors in deviation from the perfect straight-line relationship between input pressure and output voltage is called the Static Error Band, expressed as a % of Excitation Voltage. The Dynamic Error Band is the same as the Static Error Band but measured while the pressure transducer is subjected to light vibration. In practice, the pressure transducer is frequently surrounded by vibration-producing equipment, and the Dynamic Error Band is more representative of the actual error under those circumstances.

Other than the Linearity Error, all other errors are largely inherent in the structure of the particular pressure transducer model and depend upon the particular pressure range. In specifying accuracy of a pressure transducer it is important to differentiate between Linearity Error, Dynamic Error Band and Static Error Band. The type of error specified should be determined by the needs of the actual application.

The ideal output variation can be represented by a straight-line relationship between input pressure and output voltage. The various deviations from this in an actual transducer are shown in Fig. 5. With regard to establishing the error tolerances in any particular application of the transducer, the reference straight line can be defined in four ways:

  • Independent Linearity: the maximum difference between plotted points of voltage output with increasing input pressure and the most favorable straight line drawn through the points. In Sensor Systems LLC transducers, at the zero pressure input value this best straight line will pass through the output voltage point between 4% and 6%, and at the 100% input pressure value will pass through the output voltage point between 94% and 96%. Each transducer has its own individual best straight-line reference scale factor.
     
  • Fixed Gradient Linearity: same as Independent Linearity except the reference line at the zero pressure input value passes through the 5% output voltage point, and at the 100% input pressure value passes through the 95% output voltage point. The scale factor of transducer output (No Excitation Voltage/Unit Input Pressure) is the same for all transducers having this type of linearity.
     
  • Zero-Based Linearity: same as Independent Linearity except the reference line at the zero pressure value passes through the 0% output voltage point. Each transducer having this type of linearity has the same value of output (within the linearity error) at Zero input pressure, but different scale factors.
     
  • Terminal Linearity: same as Zero-Based Linearity except the reference line at the 100% input pressure value passes through the 100% output voltage point: the error band will lie within the specified value only over output voltage values between 5% and 95%. The scale factor of transducer output is the same for all transducers having this type of linearity.

MEASURING PRESSURE TRANSDUCER ACCURACY

The magnitude of individual errors making up the output Static Error Band can be determined by applying to the individual pressure transducer a repeated series of known input pressures in both an increasing and a decreasing direction, and recording the output voltages at each known pressure value; output voltages should be recorded both before and after light tapping on the transducer. Mean values of voltage output at each pressure point for the repeated runs can be calculated. The Friction Error at any pressure input is the difference between a tapped and an untapped value of voltage output. The Repeatability Error at any pressure input reached in the same direction is the difference between the mean value and the extreme value in the repeated series of tapped output voltages. The Hysteresis error at any pressure input is the difference between the mean values of tapped output voltage when measured in the increasing and decreasing pressure directions. The Linearity Error at any pressure input is the difference between the mean value of tapped output measured in the increasing pressure direction and the selected reference line value at that pressure. The Resolution Error at any pressure input is the smallest change in pressure input which produces an observable change in tapped voltage output, and can be measured during the course of applying the repeated series known input pressures. Static Error Band is obtained by plotting the untapped increasing and decreasing pressure values of output voltage for a single series of known input pressures, and observing the maximum deviation from the straight reference line indicated by the type of linearity.
 

 Using Pressure Transducers

In addition to the normal specifications such as pressure range, media, environments, etc. used to define a particular transducer for its application, other items relevant to the electromechanical transducer are important and are often defined by system requirements. These include excitation voltage, source loading, power handling, accuracy, output loading, and output signal configuration.

EXCITATION VOLTAGE, TERMINAL RESISTANCE POWER

The typical pressure transducer installation is shown schematically in Fig. 6. The power supply provides the Excitation Voltage E, either AC or DC, and the nominal scale factor of pressure transducer becomes Excitation Voltage/Pressure Range, volts/psi.

Fig. 6
Typical pressure transducer installation

For example, with 20VDC Excitation Voltage, the output scale factor of a 0-20 psi pressure transducer is nominally 1 VDC/psi. When the nominal scale factor has been decided upon, the Terminal resistance RT of the transducer can be specified. From the power supply point of view only, two considerations enter into the specification of Terminal resistance. First, there is a practical limit to the ability of the resistance element in the transducer to dissipate the heat produced by the electrical energy or power fed to it by the power supply. This Power Rating is listed in the specification table of each transducer model, typically 0.2 watts. The Power applied by the power supply can be calculated from the equation P=EI=E2/RT, and must not exceed the Power Rating of the transducer to avoid damage to it. Using 2OVDC Excitation on a transducer rated at 0.2 watts, the minimum value of Terminal Resistance is calculated to be 2000 ohms. Such a resistance will draw 0.01 amperes of current from the power supply, and the second consideration is whether the power supply can deliver this current. Clearly, for a given Excitation Voltage (a given scale factor), the higher the Terminal Resistance, the cooler the transducer resistance element will run and the lower will be the current demands from the power supply. From the point of view of the input power, then, the Terminal Resistance should be specified as high as possible.

TERMINAL RESISTANCE, LOADING LINEARITY

While the highest Terminal Resistance improves the input power situation, the considerations of electrical loading and linearity call for the lowest Terminal Resistance j the final selection of Terminal Resistance actually becomes therefore, a compromise between these opposing considerations. Electrical loading comes about in the Detector resistance RL (or other following electronic equipment fed by the moveable contact of the pressure transducer) that acts as a shunt on the transducer resistance element and thereby distorts the output of the transducer to an extent determined by the ratio RT/RL.

Fig. 7
Typical pressure transducer control circuit
 

Fig. 8
Typical pressure transducer circuit

Referring to the section "Electromechanical Transducer Performance Characteristics", above, it can be seen that for RT/RL ratios less than 0.005, there is no significant effect of electrical loading on linearity of the output. Thus, for example, when the effective electrical load resistance RL is 100,000 ohms, any Terminal Resistance less than 500 ohms will result in virtually no deterioration in output linearity due to electrical loading: if the RL value were 20,000 ohms, any greater Terminal Resistance than 100 ohms will introduce a significant linearity error. In the prior section it was determined in an example that the power rating indicates a minimum Terminal Resistance of 2000 ohms when the example transducer is excited by 20 VDC: any loading less than 400,000 ohms will produce a significant linearity error in the output.

Should it be that the power rating consideration forces the loading ratio RT/RL to exceed 0.005, then the expected maximum loading error on linearity can be read off the chart shown in Fig. 4. By compensation of the transducer calibration during manufacture these loading errors can be substantially reduced, depending upon the actual application. Where the combination of unfavorable load and high accuracy prevail, provision must be made for reducing the loading ratio, in the manner discussed.

OUTPUT VOLTAGE AND CURRENT

When the resistance of the electrical load is infinite in value, no current flows from the pressure transducer moveable contact, and the transducer output is purely voltage. This occurs in practice where, for example, the pressure transducer resistance element forms two legs of an electrical bridge network; perhaps the other two legs are made up of a manually set variable resistor used to establish a control set point, as in Fig. 7. When the voltage output of the pressure transducer exactly matches the voltage output of the hand set resistor, at the null point, no current flows from the transducer moveable contact so that the effective load resistance is infinite value.

When off null in the above bridge circuit or when, for example, the pressure transducer feeds the coil of a chart recorder pen motor, as in the schematic circuit of Fig. 8, current is drawn through the transducer moveable contact. In the null-seeking bridge circuit, the effect of being off null is to produce a change in effective local scale factor due to the linearity distortion introduced by the electrical loading of the circuit; as null is approached, however, this distortion is reduced and ultimately disappears. In the case of the chart recorder circuit, current must be drawn continuously to drive the coil against the return spring: in fact, the pen deflection is a direct measure of the current in the coil, not the voltage, and for the recorder to faithfully indicate the level of input pressure to the transducer, the transducer must be capable of delivering relatively large currents. The effective electrical loading here is equivalent to the coil resistance of the pen motor, frequently only several hundred ohms. The effect of electrical loading under these circumstances is great.

To perform accurately under circumstances of unfavorable electrical loading, it is necessary to interpose an electrical isolation amplifier between the pressure transducer and the electrical load. Such an amplifier has the characteristics of high input impedance so that the transducer feeds the equivalent of a high effective load resistance, typically 1 Megohm, and a low output impedance so that the electrical load, be it pen motor coil, relay or any other current-driven device, is fed by a source of relatively unlimited current. The Model 4105 P/l Transmitter not shown in this catalog is a self-contained device consisting of pressure transducer, isolation amplifier power supply, which is intended to provide superior performance at low-cost in such applications.

By use of modern microcircuit techniques offering precision with high reliability, isolation from burdensome electrical loads can be provided in practically every model of transducer shown in this catalog with performance characteristics to suit your particular application. Further, such internal electronic circuitry can also be incorporated to enable the generation of pressure transducer outputs proportional to Mach number, true airspeed and the like.

PRACTICAL CONSIDERATION AND PRECAUTIONS

Fig. 9
Circuits capable of damaging
the pressure transducer

Drawing current through the transducer moveable contact as a result of electrical loading produces distortions in the output voltage, but drawing excessive current may also permanently damage the transducer. Such excessive currents may be typically drawn in two ways when using the transducer. First, in testing or installing the pressure transducer, an ohmmeter may be connected between the moveable contact terminal and the low-end terminal of the transducer variable resistance, as in A of Fig. 9. With no pressure applied when the wiper is close to the end terminal or when input pressure is applied in such a way as to drive the moveable contact towards the low-end terminal, the current through the moveable contact increases, to a maximum when the wiper is directly opposite this terminal. Depending upon the ohmmeter internal voltage and resistance, relatively high currents may be drawn, damaging the resistance element area close to the low-end terminal. A similar result may be obtained when any form of circuit testing equipment containing its own power supply, such a circuit continuity testers, is connected to the moveable contact terminal of the pressure transducer. Second, in testing or installing the transducer, the Excitation Voltage may be inadvertently connected between the moveable contact terminal, as in B of Fig. 9. Obviously, the full excitation current will pass through the moveable contact and damage due to excessive current flow or burnout of the resistance element will result. Such a condition of Excitation Voltage connected to the moveable contact terminal may be artificially or momentarily introduced by circuit anomalies lying outside the pressure transducer.

For example, relay-connected power supplies may incorporate a circuit logic which produces such a misconnection under conditions of non-simultaneous engagement or relay chatter. Electrical grounding interruptions may produce similar circuit anomalies.

Analogous to electrical overload producing damage is the effect of input overpressure. All pressure transducers shown in this catalog nominally provide for overpressures 20% greater than the specified pressure range. The general effect of overpressure beyond the allowance is usually to strain the pressure-sensing element beyond its stable limit and at the same time drive the electromechanical transducer beyond its range. The effect on transducer output is usually the destruction of its accuracy, if not the complete disappearance of its output. A greater overpressure factor can be provided, to suit, within the limitations of dimensions and materials available. The danger of inadvertent overload is greatest in differential pressure transducers where the reversal of Hi and Lo inputs may occur by accident or failure further up the line. Likewise, in differential units, the absence of one pressure in the face of the application of the other may produce an overpressure situation. Safety provisions should be made in all pressure circuitry to minimize the possibility of overpressures.

Certain miscellaneous precautions should be observed in installing pressure transducers. Pressure transducers are precision devices and rough handling should be avoided. A 30" fall (from a bench top, for example) to a concrete floor produces a shock equivalent to 1000G’s, sufficient to destroy the accuracy of a transducer. Likewise, the use of the wrench flats provided on the pressure fittings when making pressure connection are preferred to wrenching on the case or base, which may produce case leakage. A controlled source of tapping which will simulate a vibration environment without damage to the transducer is the common 60-cycle bell buzzer mounted adjacent to the transducer.

To insure reliability of pressure transducer test results, the accuracy of measuring equipment should be at least ten times greater than that of the test specimen. In making absolute pressure measurements (such as in altitude transducers) a high reference vacuum in the measuring equipment must be maintained to avoid an offset in the calibration. Accidental pressure overload damage can be avoided by inserting a metering valve between the transducer and the test pressure or vacuum source. Care must be taken (especially during immersion-type leak tests) to prevent the inadvertent entry of alien fluids into the pressure transducer case volume.
 

 Environmental Considerations

VIBRATION: As a general rule, when Transducers are subjected to vibration levels above 500 Hz, a vibration isolator is required to maintain operational accuracy. Vibration levels less than 500 Hz may not require such damping depending on the model, range and G level.

In all cases, Sensor Systems LLC Application Engineering Department should be consulted prior to selection, where vibration environments are a factor.

SHOCK: The ability to withstand shock stresses is dependent on the model, range, shock profile and G level. However, most models all withstand shock levels in the order of 10 g's sinusoidal, 11 ms. without loss of continuity. Many models can withstand higher levels depending on the application. Sensor Systems LLC Applications Engineering Department will assist you in your selection for shock environments.

ACCELERATION AND POSITION ERROR: A function of range, typical values are shown for each model on the applicable data sheet.

MEDIA: In general, any media compatible with Ni-Span-C, aluminum, and stainless steel are well tolerated by
Sensor Systems LLC Transducers. Other medias can be accepted with material modifications.
 

 Special Electrical Characteristics

PRESSURE TRANSDUCERS: To meet requirements of size, performance, environment and the like, special Transducers and/or custom modification of standard models can be manufactured. Practically any combination of special requirements can be met due to the versatility of the film resistance element.

LINEARITY OR CONFORMITY: Unless otherwise indicated, the linearity shown on the catalog sheets is Independent Linearity on a voltage ratio basis. Terminal, zero based and other linearity or conformity types are available to your order.

MATCHING FUNCTIONS (Tracking and/or Phasing): Dual potentiometer outputs are available in the same case size for most models. Unless otherwise indicated both elements in a Dual unit are phased for the simultaneous output of 50% of excitation voltage to an accuracy equal to the Linearity tolerance. Simultaneous conformity throughout pressure range or other output relationships are available to suit.

NON-LINEAR OUTPUTS: A wide variety of non-linear outputs are available for applications such as airspeed, flow and altitude measurements. In addition, non-continuous outputs in the form of switches can be supplied. Typical non-linear output functions are shown in these diagrams. It is important to note that infinite resolution is an inherent property of the film-resistance element and does not vary with functional outputs.

LOAD COMPENSATION: Where desired, the transducer can be compensated during manufacture for the errors introduced by electrical loading of the wiper, thereby eliminating such errors from the system. If loading Ratio (Transducer Total Resistance) is more than .005, (Load Impedance) Sensor Systems, LLC should be contacted.

WATTAGE: The power dissipation shown on the individual specification page is the unit's rated power under ambient conditions (25°C). All units will dissipate rated power to 85°C, then derate linearly to zero power at 120°C. Special high power or high temperature units available to suit.
 

 Definitions

1. STATIC ERROR BAND represents the maximum deviation from the best straight line drawn through the coordinates of 0% pressure range, 5% +/- 1% output voltage ratio, and 100% pressure range, 95% +/- 1% output voltage ratio. This band includes the effects of linearity, friction, hysteresis, resolution, and repeatability, and is expressed as a percentage of Excitation Voltage.

2. DYNAMIC ERROR BAND is the same as Static Error Band but with friction eliminated by light vibration.

3. LINEARITY is the maximum deviation from the best straight line drawn through 5-95% +/- 1% end points taken on increasing pressure readings at a constant ambient temperature with friction error eliminated, and is expressed as a percentage of full scale excitation.

4. FRICTION ERROR is the difference in output before and after tapping or vibrating the transducer.

5. HYSTERESIS is the difference in the value of the average tapped output measured for an input of pressure, increasing from the minimum pressure, and the average tapped output, decreasing from the maximum pressure, for repeated tests.

6. RESOLUTION is the minimum increment of pressure necessary to cause a change in tapped output.

7. REPEATABILITY is the difference in value of tapped output at a single point and the average value of tapped for repeated input pressures when approached from the same direction.

NOTE:
Insulation Resistance: 50 Megohms minimum at 500 VDC.
Dielectric Strength: 750 VAC.

 

Pressure Transducers Tutorial


 

 
    

Website Design Virext Technologies                        ©2008 Sensor Systems